The NITC Inhalation Exposure System is manufactured by Korean company HCT. Very few information about toxicity is known to public and no formal studies. Requirement for toxicity studies are: Monitoring particle concentration, size and size distribution chamber.
Feaures.
- Nose-only exposure chamber system.
- Inhalation toxicity of drug, pesticides and chemicals.
- Multifunctional system can be available depending upon purpose, particles and laboratory animals.
- Nanoparticles generation/monitoring system (ISO 10801, 10808).
- Automatic generation/monitoring system.
- Systems comply with GLP, REACH, GHS and OECD guidelines.
Working Principle.
Specifications.
Particle Generation System:
- Particle size: 1nm - 100µm.
- Particle concentration:0–1012#/cm3(<300 mg/cm3)
- Particle type:NaCL, KCl, Au, Pb, Mn, SiO2, TiO2, Fe2O3, Soot, CNT, Powder.
- Enable to generate monodisperse or polydisperse aerosol
- Standard deviation of generated particle: > 1.05
- Aerosol cyclone: 5 µm cut-off.
- Adapted to ISO 10801.
Nose-only Exposure Chamber:
- Top-down flow type chamber.
- Dimension
- 1000 x 1000 x 2000 mm
- Material: SUS 316
- Inside Illumination: 200 Lux.
- Port number of chamber: 36ea (6 x 6).
- Sampling flow-rate per port: 1L/min.
- Dynamic air flow: 12 – 15 air changes per hour.
- Adapted to OECD guideline (TG-403, 412 and 413).
See more specifications on the equipment brochure.
Control & Monitoring Program.
- Environmental monitoring: Temperature, Humidity, Differential pressure, Air flow-meter, O2
- Concentration and particle size distribution.
- The state of exposure conditions.
- Inter-lock monitoring, water flushing monitoring and etc.
- Particle generation, dilution ratio and flowrate control.
- Mass flow controller.
- Make-up air valve and exposure air valve control.
- Adapted to FDA and OECD guideline.